The leak detection determination for DOP high-efficiency liquid filter is based on the core standard of **≤0.01% penetration rate (leakage rate)**. It is accompanied by repair and re-test rules, which need to cover the filter material, seal, and frame, and meet the specifications such as ISO 14644-3 and GB/T 6165-2008. Here are the determination criteria and practical key points:
Core determination indicators
Penetration rate threshold: The continuous and stable penetration rate of all scanable points of the filter ≤ 0.01% (that is, one ten-thousandth of the upstream concentration), any point exceeding this threshold is judged as不合格, and it needs to be marked for repair or replacement.
Scan coverage range: Filter material surface, filter material – frame bonding area, frame – liquid tank sealing surface, liquid tank – static pressure box / enclosure joint, frame joint seam, etc.
Detection parameters: Scan speed ≤ 5cm/s, probe distance from the filter surface is approximately 25mm, line overlap ≥ 10mm; the upstream aerosol concentration is stable at 10–20μg/mL (or meets the instrument linear range).
Repair and re-test rules
| Project | Qualified upper limit | Re-test requirements |
| Single repair length | ≤38mm | After repair, a re-measurement of the entire area was conducted, and the penetration rate remained ≤ 0.01% |
| Cumulative repair area | ≤ 3% of the filter material area | If it exceeds the limit, it must be replaced entirely. |
| Repair method | Special sealant / patch, compacted without air bubbles | Must not affect the sealing of the liquid tank and the uniformity of the airflow |
Qualification determination process
Upstream dust emission and concentration calibration: Confirm the stability of the upstream concentration, and perform zero adjustment and linear calibration on the spectrophotometer.
Zone scanning and data recording: Scan each zone, record the penetration rate at each point, and mark and locate the abnormal points.
Result determination: If the penetration rate of all points is ≤ 0.01% and the repair is compliant, it is determined as qualified; otherwise, it needs to be repaired/ replaced and retested until qualified.
Report and traceability: Record the equipment number, detection date, upstream concentration, maximum penetration rate, repair location / area, retest results, etc., and archive for retention.
Key standards basis
ISO 14644‑3: The maximum penetration rate of scanable high-efficiency filters is ≤ 0.01%.
GB/T 6165‑2008: The on-site leak detection penetration rate of high-efficiency filters (0.3μm) is ≤ 0.01%.
FDA/GMP: The pharmaceutical industry requires that filters and seals have no leakage and the penetration rate is ≤ 0.01%.